Description
Base Unit Specifications | |
Temperature range at 23°C | 33°C to 350°C (91°F to 662°F) |
Display accuracy | ±0.2°C Full Range |
Stability | ±0.02°C at 33°C ±0.02°C at 200°C ±0.03°C at 350°C |
Axial uniformity at 40 mm (1.6 in) | ±0.04°C at 33°C ±0.1°C at 200°C ±0.2°C at 350°C |
Radial uniformity | ±0.01°C at 33°C ±0.015°C at 200°C ±0.02°C at 350°C |
Loading effect (with a 6.35 mm reference probe and three 6.35 mm probes) | ±0.015°C Full Range |
Hysteresis | 0.03 |
Operating conditions | 0°C to 50°C, 0% to 90% RH (non-condensing) |
Environmental conditions (for all specifications except temperature range) | 13°C to 33°C |
Immersion (well) depth | 150 mm (5.9 in) |
Insert OD | 25.3 mm (1.00 in) |
Heating time | 5 min: 33°C to 350°C |
Cooling time | 32 min: 350°C to 33°C 14 min: 350°C to 100°C |
Resolution | 0.01° |
Display | LCD, °C or °F user-selectable |
Size (H x W x D) | 290 x 185 x 295 mm (11.4 x 7.3 x 11.6 in) |
Weight | 7.3 kg (16 lb) |
Power requirements | 100 V to 115 V (±10%), 50/60 Hz, 1400 W 230 V (±10%), 50/60 Hz, 1800 W |
Computer interface | RS-232 and 9930 Interface-it control software included |
-P specifications | |
Built-in reference thermometer readout accuracy (4-Wire reference probe)1 | ±0.010°C at -25°C ±0.015°C at 0°C ±0.020°C at 50°C ±0.025°C at 150°C ±0.030°C at 200°C ±0.040°C at 350°C ±0.050°C at 420°C ±0.070°C at 660°C |
Reference resistance range | 0 ohms to 400 ohms |
Reference resistance accuracy2 | 0 ohms to 42 ohms: ±0.0025 ohms 42 ohms to 400 ohms: ±60 ppm of reading |
Reference characterizations | ITS-90, CVD, IEC-751, Resistance |
Reference measurement capability | 4-wire |
Reference probe connection | 6-pin Din with Infocon Technology |
Built-in RTD thermometer readout accuracy | NI-120: ±0.015°C at 0°C PT-100 (385): ±0.02°C at 0°C PT-100 (3926): ±0.02°C at 0°C PT-100 (JIS): ±0.02°C at 0°C |
RTD resistance range | 0 ohms to 400 ohms |
RTD resistance accuracy‡ | 0 ohms to 25 ohms: ±0.002 ohms 25 ohms to 400 ohms: ±80 ppm of reading |
RTD characterizations | PT-100 (385),(JIS),(3926), NI-120, Resistance |
RTD measurement capability | 4-wire RTD (2-,3-wire RTD w Jumpers only) |
RTD connection | 4 terminal input |
Built-in TC thermometer readout accuracy | Type J: ±0.7°C at 660°C Type K: ±0.8°C at 660°C Type T: ±0.8°C at 400°C Type E: ±0.7°C at 660°C Type R: ±1.4°C at 660°C Type S: ±1.5°C at 660°C Type M: ±1.4°C at 660°C Type L: ±0.7°C at 660°C Type U: ±0.75°C at 600°C Type N: ±0.9°C at 660°C Type C: ±1.1°C at 660°C |
TC millivolt range | –10 mV to 75 mV |
Voltage accuracy | 0.025% of reading + 0.01 mV |
Internal cold junction compensation accuracy | ±0.35°C (ambient of 13°C to 33°C) |
TC connection | Small connectors |
Built-in mA readout accuracy | 0.02% of reading + 2 mV |
mA range | Cal 4-22 mA, Spec 4-24 mA |
mA connection | 2 terminal input |
Loop power function | 24 V DC loop power |
Built-in electronics temperature coefficient (0°C to 13°C, 33°C to 50°C) | ±0.005% of range per °C |